Customer need
Maintenance and survey work require a measurement method and range suitable for the gas encountered.
NOVEL · Product and System Solutions
Gas Detectors and Industrial Measurement Instruments
New Cosmos XP–3300II series portable gas detectors have been developed to measure concentrations of flammable gases, solvent vapours, and hazardous volatile components with high sensitivity, and to detect ppm-level leaks at an early stage. Thanks to their compact and ergonomic design, they offer easy portability during field work and allow comfortable single-handed operation.
This series is actively used especially by municipal gas distribution utilities, maintenance, and emergency response teams for gas leak detection, line checking, and risk assessment tasks. In environments with a risk of gas accumulation, it instantaneously detects gas concentrations reaching hazardous levels and warns the user via audible, visual, and vibration alarms. In this way, it helps prevent explosion, fire, and poisoning risks.
From need to solution
New Cosmos XP–3310II, XP-3340II, and XP-3360II Series: selection starts with the measurement or system output your team needs for its decision.
Maintenance and survey work require a measurement method and range suitable for the gas encountered.
Readings and alarm indications from the fitted sensors help assess conditions at the inspection point.
Plan screening, closer investigation and repeat measurements, with pre-use checks and site procedures as part of the measurement process.
The XP–3300II series is an ideal solution for enhancing personnel safety in working environments with restricted ventilation, such as underground construction, tunnels, sewer lines, confined spaces, and industrial facilities. Thanks to its durable housing construction and field-suitable design, it delivers reliable performance under demanding environmental conditions.
The multi-gas models in the series combine oxygen and combustible gas detection in a single instrument, providing the user with more comprehensive safety monitoring. With a wide range of sensor options for various gas types, it adapts to diverse application areas and supports businesses' occupational health and safety standards.
Select the target combustible gas and calibration when ordering. XP-3310II: 0–100 %LEL; XP-3360II: 0–5000 or 0–10000 ppm; XP-3360II-W: 0–100 %LEL with ppm display; XP-3340II: 0–100 %vol. Gas and range depend on the model.
| Target Gas | Flammable Gases, Solvent Vapours |
|---|---|
| Detection Method | Catalytic |
| Measurement Range | 0.0–100.0%LEL (100.1–110.0%LEL) |
| Reference Indication | Activate for indication in ppm. |
| Resolution | 0.1%LEL or 1ppm (Dual Scale!) |
| Preset Alarm Values | 20%LEL |
| Response Time | T90: under 30 seconds |
| Continuous Operating Time | Approximately 15 hours |
| Sampling Method | Suction Type with Built-in Pump |
| Alarm Indications | Gas Alarm: Alarm icon on built-in screen, audible and red visual emergency warning Device Alarm: Error icon on built-in screen, yellow visual status warning |
| Power Supply | Rechargeable nickel-metal hydride AA battery x4: New Cosmos HR-3UTG, can be operated with optional alkaline AA batteries |
| Operating Temperature and Humidity | -20 to +50oC 95%RH non-condensing |
| Compliance | ATEX: Ex II 1G Ex ia da IIC T4 Ga1 IECEx: Ex ia da IIC T4 Ga1 CE (ATEX, EMC, RoHS, RE and LVD directives) |
| Protection Class | IP67 |
| Features | Self-diagnostics (sensor error), zero adjustment, battery level indication, flashlight, peak hold display, backlit LCD display, muting of audible alarm during gas alarm, clock display, alarm test, alarm tone selection, volume reduction |
| Dimensions and Weight | 91 x 164 x 44mm and 460g (including batteries) |
| Accessories | Gas sampling hose (1m), shoulder strap, drain filter and filter elements, AA alkaline battery (or rechargeable nickel-metal hydride AA battery) x4, screen protector film x3 |
| Target Gas | Combustible Gases Solvent Vapour |
|---|---|
| Detection Method | Catalytic |
| Measurement Range | 0–5000ppm or |
| Reference Indication | 0 – 10000 ppm (5001–5500ppm or 10001–11000ppm) |
| Accuracy | ≤10%F.S.: ±2.5%F.S. >10%F.S.: ±5% |
| Resolution | 1ppm |
| Preset Alarm Values | 250ppm or 500ppm |
| Response Time | T90: 30 seconds |
| Continuous Operating Time | Approximately 15 hours |
| Sampling Method | Extractive |
| Alarm Indications | Gas Alarm: Beeping sound with flashing red light and AL icon on LCD screen Device Alarm: Beeping sound with flashing yellow light and error message on LCD screen |
| Power Supply | Non-rechargeable AA battery x4: Toshiba LR6, Panasonic LR6X, Duracell MN1500, Energizer E91 or Varta 4106 Rechargeable nickel-metal hydride AA battery x4: New Cosmos HR-3UTG |
| Operating Temperature and Humidity | -20 to +50oC 95%RH non-condensing |
| Compliance | ATEX: Ex II 1G Ex ia da IIC T4 Ga1 IECEx: Ex ia da IIC T4 Ga1 CE (ATEX, EMC, RoHS, RE and LVD directives) |
| Protection Class | IP67 |
| Features | Self-diagnostics (sensor error), zero adjustment, battery level indication, flashlight, peak hold display, backlit LCD display, muting of audible alarm during gas alarm, clock display, alarm test, alarm tone selection, volume reduction |
| Dimensions and Weight | 91 x 164 x 44mm and 460g (including batteries) |
| Accessories | Gas sampling hose (1m), shoulder strap, drain filter and filter elements, AA alkaline battery (or rechargeable nickel-metal hydride AA battery) x4, screen protector film x3 |
| Target Gas | Combustible Gases Solvent Vapour |
|---|---|
| Detection Method | Catalytic |
| Measurement Range | 0–100%LEL |
| Reference Indication | 100.1 – 110.0% LEL |
| Accuracy | ±5%F.S. |
| Resolution | 0.1%LEL |
| Preset Alarm Values | 20%LEL |
| Response Time | T90: 30 seconds |
| Continuous Operating Time | Approximately 15 hours |
| Sampling Method | Extractive |
| Alarm Indications | Gas Alarm: Beeping sound with flashing red light and AL icon on LCD screen Device Alarm: Beeping sound with flashing yellow light and error message on LCD screen |
| Power Supply | Non-rechargeable AA battery x4: Toshiba LR6, Panasonic LR6X, Duracell MN1500, Energizer E91 or Varta 4106 Rechargeable nickel-metal hydride AA battery x4: New Cosmos HR-3UTG |
| Operating Temperature and Humidity | -20 to +50oC 95%RH non-condensing |
| Compliance | ATEX: Ex II 1G Ex ia da IIC T4 Ga1 IECEx: Ex ia da IIC T4 Ga1 CE (ATEX, EMC, RoHS, RE and LVD directives) |
| Protection Class | IP67 |
| Features | Self-diagnostics (sensor error), zero adjustment, battery level indication, flashlight, peak hold display, backlit LCD display, muting of audible alarm during gas alarm, clock display, alarm test, alarm tone selection, volume reduction |
| Dimensions and Weight | 91 x 164 x 44mm and 460g (including batteries) |
| Accessories | Gas sampling hose (1m), shoulder strap, drain filter and filter elements, AA alkaline battery (or rechargeable nickel-metal hydride AA battery) x4, screen protector film x3 |
With its concentration conversion function for 32 different gas types, it enables reliable monitoring of different gas components via a single instrument.
With a housing construction resistant to demanding field conditions, it offers reliable operation in industrial environments, confined spaces, and outdoor applications.
With model options featuring different measuring ranges, it covers a wide application spectrum from low-concentration leak detection to high-density gas measurements.
The ability to select up to five target gases provides the user with flexible measurement capability tailored to the site and application.
With alkaline battery and rechargeable internal battery options, it supports flexible power source use for prolonged and uninterrupted field operations.
Thanks to the timer function, measurement periods can be controlled and periodic field inspections are simplified.
The built-in flashlight function assists in working and measuring safely in low-light environments.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.
Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.