Sensidyne GK2.69 Thoracic / Respirable Cyclone

Select thoracic sampling at 1.6 L/min or respirable sampling at 4.2 L/min with one 37 mm cyclone head.

Sensidyne GK2.69 Thoracic / Respirable Cyclone

Choose the Fraction Before Sampling

GK2.69 separates particles by aerodynamic size before collecting the selected fraction on a filter. The required fraction is selected by flow rate; the same sample does not separately quantify both fractions. A compatible Gilian or other personal sampling pump must maintain the selected flow against the filter pressure drop.

Request a Quote from NOVEL ↗

From need to solution

Sample the required aerosol fraction

Sensidyne GK2.69 Thoracic / Respirable Cyclone: selection starts with the measurement or system output your team needs for its decision.

01

Customer need

Respirable or thoracic sampling needs an appropriate size-selective head and specified flow, in addition to a filter.

02

Product output

At the correct flow, the cyclone supports collection of the intended fraction on the filter.

03

Decision supported

Select the head, pump and filter together and check the assembled train; thoracic and respirable settings are not interchangeable.

Why Use It?

Workplace exposure assessment can require the thoracic or respirable fraction rather than undifferentiated total dust. The filter and analytical method are selected for the substance; gravimetric dust mass alone does not identify crystalline silica or other specific constituents.

Setup and Flow Verification

Fit the filter-containing section of the three-piece cassette to the cyclone O-ring seat. Connect the cassette outlet to the pump, secure the head at the lapel and route the hose over the shoulder. Calibrate with a fresh filter and the complete train. Select 1.6 or 4.2 L/min for the intended fraction; consult the manual for connections, cleaning and pressure requirements.

Pump and Filter Compatibility

The manufacturer reports 11.5 cm H₂O pressure drop with a 37 mm GF/A glass-fibre filter at 4.2 L/min, and 3.4 cm H₂O at 1.6 L/min. A 0.8 µm MCE filter at 1.6 L/min is listed at 11.9 cm H₂O. These are specified filter/flow combinations, not a universal pressure drop for every membrane.

Air sampling setup video
Sampling setup example. Follow the manual for the selected cyclone and cassette.

Technical Specifications

Model and Sampling Setup
Manufacturer code
811-9926-01
Cassette
37 mm, three-piece
Thoracic sampling
1.6 L/min · D50: 10 µm
Respirable sampling
4.2 L/min · D50: 4 µm
Flow calibration
Complete cyclone + new filter + pump train
Sampling result
Filter sample for subsequent laboratory analysis

Accessories and Consumables

37 mm three-piece filter cassette, method-compatible membrane, cassette-to-pump hose with Luer connection, suitable flow calibrator and carrying clip. Confirm the ordered kit; the presence of an item in this list does not imply standard delivery. Browse Filter and Cassette Codes →

Frequently Asked Questions

What information is needed to select Sensidyne GK2.69 Thoracic / Respirable Cyclone?

Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.

Where can I find accessories and technical documents?

Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.

Manufacturer Source

Sensidyne GK2.69 · 811-9926-01 ↗

NOVEL Technical Support

Share your application, target parameter and required output. We can help review the model and accessory selection.

Request a Quotation from NOVEL ↗