Fit the filter
Use gloves and clean tweezers. Open the cassette, seat the method-specified 25 mm filter without damaging it, close the cassette securely and install it in the head.
NOVEL · Technical Expertise and Company Information
Occupational Safety and Environmental Measuring Instruments

Collect the inhalable fraction in the worker’s breathing zone for laboratory analysis. Pair the reusable head and removable 25 mm cassette with a pump set to 2 L/min; select filter material and analysis with your laboratory.
Request a Quote from NOVEL ↗The head and replaceable cassette form a sampling system: send the exposed cassette to the laboratory, then fit another cassette to continue work. The inhalable convention has approximately 50% sampling efficiency at 100 µm; this is not a sharp cutoff that captures every particle below 100 µm. Inhalable and respirable fractions are different.
ZEFON states that the sampler is manufactured under licence from the Institute of Occupational Medicine; I.O.M. is a registered trademark, number 2043339E. Its use within MDHS 14/4 requires the complete sampling and analytical procedure, rather than a blanket claim of compliance with all standards.
Use gloves and clean tweezers. Open the cassette, seat the method-specified 25 mm filter without damaging it, close the cassette securely and install it in the head.
Connect head, tubing and pump. Use adapter ZA0307 or the manufacturer-approved calibration chamber; verify 2 L/min before and after sampling.
Place the inlet in the breathing zone, follow method duration and volume, then cap the cassette, label it and send it to the laboratory. Account for wall deposits as required by the method.
Check seals and cassette fit. The manufacturer specifies water for cleaning O-rings; organic solvents can damage them. Follow the manual for the remaining components.
| Code | Description | Filter / Pack |
|---|---|---|
| ZA0300 | Conductive plastic head, plastic cassette | 25 mm · 2 L/min |
| ZA0303 | Conductive plastic head, stainless-steel cassette | 25 mm · 2 L/min |
| ZA0301 | Reusable plastic cassette and transport clip | 25 mm |
| ZA0304 | Stainless-steel replacement cassette | 25 mm |
| ZA0307 | Calibration adapter | IOM |
| ZA0305 | O-ring pack | 3 |
| CLCL-DIS0810 | Disposable sampler with MCE capsule | 10 |
| PVPV-DIS5010 | Disposable sampler with PVC capsule | 10 |
| PVPV-DIS5010P | Preweighed PVC version | 10 |
| CLCL-DIS0810-F | Foam-equipped version | 10 |
| PVPV-DIS5010U | Unassembled PVC version | 10 |


Filter material, pore size, weighing procedure and analysis must follow the target method. A matching diameter alone does not establish suitability.
A cyclone targets a specified size-selective fraction. The inhalable IOM system and a respirable cyclone are not interchangeable.