Customer need
Gas-quality control and process investigation require an H₂S range and sampling conditions suited to the task.
NOVEL · Product and System Solutions
Gas Detectors and Industrial Measurement Instruments

From need to solution
Analytical Industries Inc (AII) GPR Industrial Series: selection starts with the measurement or system output your team needs for its decision.
Gas-quality control and process investigation require an H₂S range and sampling conditions suited to the task.
The selected analyser configuration measures H₂S concentration in the gas sample.
Support gas-treatment and quality-control decisions, checking sample lines, gas composition and sensor suitability together.
| Detected Gas Range | 0-20, 0-50, 0-100 ppm |
|---|---|
| Connection Outputs | 0-1V |
| Power Supply | Rechargeable battery, Operating time ~8h |
| Dimensions and Weight | 910 x 149 x 40 mm - 3.6kg |
| Safety Classification | ATEX: II 2 G Ex ib IIB T4 Gb Tamb +5°C to +45°C |
| Measurement Accuracy | < ±2% |
| Response Time | T90: < 60s |
| Sensitivity | 0.1ppm |
| Sensor Life | ~2 years (at 25°C temperature and 1atm pressure) |
| Display | Backlit LCD Display |
| Display Resolution | 0.1ppm |
| Operating Temperature | -10°C to +45°C temperature |
| Detected Gas Range | 0-20, 0-50, 0-100 ppm (Standard Ranges) 0-500, 0-1000, 0-2000 ppm (Optional Ranges) |
|---|---|
| Connection Outputs | 4-20mA and 1-5V DC |
| Power Supply | 110-220 V AC |
| Dimensions and Weight | 69 x 104 x 34 mm - 10kg |
| Alarm | Adjustable 2-stage alarm |
| Safety Classification | ATEX: II 2 G Ex ib IIB T4 Gb Tamb +5°C to +45°C UL: United States: UL 1203, UL 913, UL 508 |
| Measurement Accuracy | < ±2% |
| Response Time | T90: < 60s |
| Sensitivity | 0.1ppm |
| Sensor Life | ~2 years (at 25°C temperature and 1atm pressure) |
| Display | Backlit LCD Display |
| Display Resolution | 0.1ppm |
| Operating Temperature | -10°C to +45°C temperature |
Low Power Consumption Electrochemical Sensor Technology
Precise Measurement
ATEX Certified (Suitable for Potentially Explosive Atmospheres)
Sample Dilution with Self-Aspirating Feature
Optional Liquid Drain System
Sensor Life Approximately 24 Months
Easy Operation and Calibration
High-Quality Stainless Steel Internal Components
Sensor Technology
AII hydrogen sulphide sensors are manufactured to be compact and highly sensitive. Additionally, their power consumption is very low compared to equivalents. The low power consumption provides a perfect match for the GPR-7100 portable hydrogen sulphide analyser, allowing the instrument to operate longer on a single charge. As with almost all toxic gas sensors, AII hydrogen sulphide sensors require a small amount of oxygen to operate.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.
Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.