NOVEL · Product and System Solutions

Gilian 12

Occupational Safety and Environmental Measuring Instruments

Gilian 12

Heavy-Duty Dust, Gas, and Air Sampling Pump (High Flow)

Request a Quote from NOVEL ↗

From need to solution

Collect field samples for laboratory analysis

Gilian 12: selection starts with the measurement or system output your team needs for its decision.

01

Customer need

Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.

02

Product output

Air passed through the appropriate filter, cassette, cyclone or sorbent tube creates a sample for analysis; the pump alone does not measure contaminant concentration.

03

Decision supported

Relate the laboratory result to sampled air volume, selecting media and flow together and calibrating the assembled sampling train.



Product Features

The Gilian 12 is a personal air sampling pump that provides a constant flow rate of 4–12 L/min with ±5% flow control. It features a removable NiMH battery with a charging time of under 4 hours, 4-stage programming, delayed start capability, and a display showing flow rate and total volume. Operational runtime depends on the operating flow rate and backpressure conditions.

Verify the flow before and after sampling with a suitable calibrator and the sampling train in place. Select the media and sampling head according to the analytical method.

Sampling and Standards Approach

ISO 13137 approach: Required flow rate stability and performance specifications for personal air sampling pumps serve as a fundamental benchmark when selecting pumps for sampling dust, aerosols, fumes, mists, fibres, gases, and vapours. Sampling flow rates and media must always be selected in accordance with the applicable national or international standard method; the pump should be calibrated and verified with an appropriate flow calibrator before and after each sampling event. Model-specific certification claims must be confirmed independently against the manufacturer's current documentation.

Technical Specifications

Show Technical Details
Flow Range 4000–12000 cc/min (4.0–12.0 L/min)
Constant Flow Control ±5% of set flow, 4–12 L/min
Display Indications Flow rate, total sampling time and volume, flow fault, battery power indicator, programming mode
Runtime Minimum 8 hours only at the flow/backpressure combinations in the manufacturer datasheet.
Weight 580 g
BatteryRemovable rechargeable NiMH
ChargingUnder 4 hours

Accessories, Spares and Kits

Order codes identify specific configurations. Kit contents, power-cord options and current availability are confirmed at quotation.

Manufacturer CodeAccessory / configuration
910-1601-EUStarter kit with Euro cord
910-1605-EUFive-pack kit with Euro cord
811-0802-02Charger with Euro power cord
783-0018-01Battery
811-1501-01Filter screen and O-ring
202783-RCharging cap

Sampling Heads, Media and Flow Calibration

Gilian 12 covers 4–12 L/min directly. GK2.69 respirable at 4.2 L/min may be used; thoracic 1.6 L/min and IOM 2 L/min are outside the direct range. Set up the complete sampling train, verify flow at the sampling-head inlet before and after use, and record the result. A field flow check does not replace periodic laboratory calibration.

Gilibrator 3 Air Flow Calibrator

Gilibrator 3 · hücre seçimi

Gilibrator 3 Air Flow Calibrator

Dry cells: low 5–450 mL/min, standard 50–5000 mL/min, high 1–30 L/min. Select the cell covering the sampling flow; check before/after sampling with head, filter and tubing assembled.

Sectors

Alternative Products

You can examine product details and technical specifications by selecting an alternative suited to your requirements.

Frequently Asked Questions

What information is needed to select Gilian 12?

Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.

Where can I find accessories and technical documents?

Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.

NOVEL technical support

For your application,
let us select the right solution together.

Request a Quote from NOVEL ↗

A draft opens in your email client with the subject header of the relevant product or category.