Customer need
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
NOVEL · Product and System Solutions
Occupational Safety and Environmental Measuring Instruments

From need to solution
Gilian 5000: selection starts with the measurement or system output your team needs for its decision.
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
Air passed through the appropriate filter, cassette, cyclone or sorbent tube creates a sample for analysis; the pump alone does not measure contaminant concentration.
Relate the laboratory result to sampled air volume, selecting media and flow together and calibrating the assembled sampling train.
| Flow Range | Total: 20–5,000 cc/min; high flow: 800–5,000 cc/min; low flow: 20–800 cc/min (with low-flow constant-pressure adapter). |
|---|---|
| Sensitivity | ± 5% (for 1000–5000 cc/min) |
| Display Indications | Flow rate, total sampling time and volume, flow fault, battery power indicator, programming mode |
| Battery Life | Min. 8 hours |
| Weight | 580 g |
| Compliance / Approvals | EN 1232 (Type P), FM, CE, Ex II 1G, Ex IIC T4, ATEX 94/9/EC, EMC 2004/108/EC (RFI), LVD 2006/95/EC |
Wide Flow Range Enabling Both Dust and Gas Sampling An adjustable flow range from 750 to 5000 cc/min accommodates all dust sampling requirements, whilst the low flow adaptor lowers the lower flow limit to 20 cc/min, enabling use for gas sampling as well.
Industry-Leading Capability in Overcoming High Back Pressure! Its advanced pneumatic design prevents clogging issues, ensuring problem-free sampling even in the most demanding dusty environments and through the finest pore filters. The pump will never let you down in the field.
User-Friendly Advanced Design Gilian brand pumps are engineered to lighten the workload, assisting site operatives with straightforward menu layouts and intuitive operation. In addition, features such as programmable run times and automatic restart are designed to provide convenience during field sampling.
Complete Control via the User Display The digital display provides clear visibility of parameters including flow rate, run time, total sampling volume, and remaining battery life.
Comprehensive Certification Compliance and Explosion-Proof Capability EN 1232 Type P compliant; features international approvals, intrinsic safety (IS), and hazardous area certification (ATEX).
Straightforward Calibration Features Flow adjustment and calibration require no dedicated adaptor or complex setup; it is compatible with all flow calibrators, and calibration is completed in seconds via the on-board keypad. Flow calibration is maintained even if ambient atmospheric conditions fluctuate after calibration.
Select the high-flow train and the required low-flow adapter/tube holder separately for Gilian 5000. Consider GK2.69 at 1.6 and 4.2 L/min and IOM at 2 L/min. Set up the complete sampling train, verify flow at the sampling-head inlet before and after use, and record the result. A field flow check does not replace periodic laboratory calibration.

GK2.69
Thoracic fraction at 1.6 L/min; respirable fraction at 4.2 L/min. Select the fraction by flow and confirm that the pump maintains it under filter load.

IOM · 2 L/dk
Samples the inhalable dust fraction at 2 L/min. Select the filter and internal cassette for the weighing/analytical method.

25 / 37 mm · metot seçimi
Choose MCE, PVC, PTFE or another filter for the analyte and laboratory method. Match cassette diameter, sections, pore size and flow in the sampling plan.

Sorbent · metot seçimi
For gas/vapour sampling select sorbent, tube dimensions, duration and low flow by method. Tube holder, breaker and connection tubing complete the train.
Impinger · sıvı tutucu
Liquid collection methods require an upright bottle holder and a trap protecting the pump. The analytical method determines flow, absorbing liquid and sample volume.

Gilibrator 3 · hücre seçimi
Dry cells: low 5–450 mL/min, standard 50–5000 mL/min, high 1–30 L/min. Select the cell covering the sampling flow; check before/after sampling with head, filter and tubing assembled.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
The Gilian 5000 can be used at high flow for dust, aerosol, mist, and fume sampling, and with the low flow adaptor for gas and vapour sampling applications.
According to the technical specifications on the page, the pump has a total flow range of 20–5000 cc/min. High flow is specified as 800–5000 cc/min, and low flow with the low flow adaptor is specified as 20–750 cc/min.
The low flow adaptor enables the pump flow rate to be reduced to lower flows suitable for gas and vapour sampling applications.
Parameters such as flow rate, total sampling time and volume, flow fault, battery status, and programming mode can be viewed on the digital display.
Flow rate verification must be performed before and after sampling using a suitable flow calibrator. According to the page specifications, the pump enables flow adjustment and calibration routines to be carried out via on-board controls.
The current page lists intrinsically safe / Ex characteristics and various compliance details, including ATEX, for the Gilian 5000. Current model certifications must be confirmed against manufacturer documentation prior to procurement or site use.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.