Customer need
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
NOVEL · Product and System Solutions
Occupational Safety and Environmental Measuring Instruments

From need to solution
Gilian 800i: selection starts with the measurement or system output your team needs for its decision.
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
Air passed through the appropriate filter, cassette, cyclone or sorbent tube creates a sample for analysis; the pump alone does not measure contaminant concentration.
Relate the laboratory result to sampled air volume, selecting media and flow together and calibrating the assembled sampling train.
| Flow Range | 200–800 cc/min (Total) |
|---|---|
| Sensitivity | ±%5 |
| Display Indications | Flow rate, total sampling time and volume, flow fault, battery power indicator, programming mode |
| Battery Life | Min. 8 hours |
| Weight | 580 g |
| Compliance / Approvals | EN 1232 (Type G), FM, CE, Ex II 1G, Ex IIC T4, ATEX 94/9/EC, EMC 2004/108/EC (RFI), LVD 2006/95/EC |
Advanced High-Performance Model for VOC Sampling With an adjustable flow range between 200 and 800 cc/min, this pump is designed for VOC sampling under demanding conditions, featuring robust back-pressure capability.
Industry-Leading Capability in Overcoming High Back Pressure! Its advanced pneumatic design prevents clogging and maintains operation in the most demanding sampling environments. It features 80" H2O back-pressure capability.
User-Friendly Advanced Design Gilian brand pumps are engineered to lighten the workload, assisting site operatives with straightforward menu layouts and intuitive operation. In addition, features such as programmable run times and automatic restart provide convenience during field sampling.
Complete Control via the User Display The digital display provides clear visibility of parameters including flow rate, run time, total sampling volume, and remaining battery life.
Comprehensive Certification Compliance and Explosion-Proof Capability EN 1232 Type G compliant; features international approvals, intrinsic safety (IS), and hazardous area certification (ATEX).
Straightforward Calibration Features Flow adjustment and calibration require no dedicated adaptor or complex setup; it is compatible with all flow calibrators, and calibration is completed in seconds via the on-board keypad. Flow calibration is maintained even if ambient atmospheric conditions fluctuate after calibration.
800i serves 200–800 mL/min low-flow trains. IOM at 2 L/min and GK2.69 thoracic/respirable heads fall outside this range. Set up the complete sampling train, verify flow at the sampling-head inlet before and after use, and record the result. A field flow check does not replace periodic laboratory calibration.

Sorbent · metot seçimi
For gas/vapour sampling select sorbent, tube dimensions, duration and low flow by method. Tube holder, breaker and connection tubing complete the train.
Impinger · sıvı tutucu
Liquid collection methods require an upright bottle holder and a trap protecting the pump. The analytical method determines flow, absorbing liquid and sample volume.

Gilibrator 3 · hücre seçimi
Dry cells: low 5–450 mL/min, standard 50–5000 mL/min, high 1–30 L/min. Select the cell covering the sampling flow; check before/after sampling with head, filter and tubing assembled.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.
Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.