Customer need
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
NOVEL · Product and System Solutions
Occupational Safety and Environmental Measuring Instruments

From need to solution
Gilian Aircon-2: selection starts with the measurement or system output your team needs for its decision.
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
Air passed through the appropriate filter, cassette, cyclone or sorbent tube creates a sample for analysis; the pump alone does not measure contaminant concentration.
Relate the laboratory result to sampled air volume, selecting media and flow together and calibrating the assembled sampling train.
| Flow Range | 2.0 L/min – 30.0 L/min (Litres/minute) |
|---|---|
| Sensitivity | ± %5 |
| Battery Life | Min. 8 hours |
| Weight | Main unit: 5.4 kg; power module: 0.9 kg; 4-hour battery pack: 5.2 kg. |
| Display / Indicators | Electronic pressure readout, full-function programming, instantaneous fault detection, programmable memory (up to 3 programs), low battery warning, rotameter indicating flow |
Extended Flow Range Designed with a 2–30 L/min flow range for particulate sampling in designated areas.
Time-Based Programmability Up to three separate time-based programs can be configured (up to 99 hours)
Touchscreen LCD Display Start-stop, pause, delay, and cycle programming operations can be performed via the touchscreen LCD display.
Instantaneous Fault Detection Function When the pump shuts down due to an error during sampling, a "FAULT" message appears on screen to alert the operator.
Real-Time Pressure Monitoring Pressure can be checked instantly on the display.
AirCon-2 is a 2–30 L/min area sampler for fixed-point cassette/media trains. Thoracic 1.6 L/min is outside range. Select a calibrator cell covering the actual high flow, up to 30 L/min. Set up the complete sampling train, verify flow at the sampling-head inlet before and after use, and record the result. A field flow check does not replace periodic laboratory calibration.

25 / 37 mm · metot seçimi
Choose MCE, PVC, PTFE or another filter for the analyte and laboratory method. Match cassette diameter, sections, pore size and flow in the sampling plan.
Impinger · sıvı tutucu
Liquid collection methods require an upright bottle holder and a trap protecting the pump. The analytical method determines flow, absorbing liquid and sample volume.

Gilibrator 3 · hücre seçimi
Dry cells: low 5–450 mL/min, standard 50–5000 mL/min, high 1–30 L/min. Select the cell covering the sampling flow; check before/after sampling with head, filter and tubing assembled.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.
Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.