Customer need
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
NOVEL · Product and System Solutions
Occupational Safety and Environmental Measuring Instruments

From need to solution
Gilian GilAir-5: selection starts with the measurement or system output your team needs for its decision.
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
Air passed through the appropriate filter, cassette, cyclone or sorbent tube creates a sample for analysis; the pump alone does not measure contaminant concentration.
Relate the laboratory result to sampled air volume, selecting media and flow together and calibrating the assembled sampling train.
| Flow Range | Total: 1–5,000 cc/min; high flow: 850–5,000 cc/min; constant low-flow module: 20–500 cc/min; constant-pressure module: 1–750 cc/min. |
|---|---|
| Sensitivity | ± 5% (for 1000–5000 cc/min) |
| Flow Fault Warning | Available |
| Battery Life | Min. 8 hours |
| Weight | 638g |
Both Dust and Gas Sampling via 1-5000cc/min Range Ideally designed for occupational safety measurements, this pump has a flow capacity between 1-5000cc/min to perform both gas and dust sampling. It requires the constant low flow module for samplings between 5-500cc/min, and the constant pressure module for samplings between 1-750cc/min.
3 Distinct GilAir-5 Models: Basic, Clock, and Program The GilAir-5 instrument has 3 distinct models: respectively, the Basic model is an instrument with standard features, the Clock model displays sampling time, whilst the Program model features auto start/stop and delayed start capabilities up to 9999 min.
User-Friendly Advanced Design Gilian brand pumps are operator-friendly in the field, with a straightforward menu design and easy operation intended to reduce workload.
Compact, Lightweight, Robust, and Economical Pump Unlike competing pumps, the Gilian GilAir-5 is an exceptionally durable, easy-to-use, practical air sampling pump. Unlike traditional pumps, it causes no strain when worn on the person due to its compact dimensions. It offers an economical alternative to premium air sampling pumps without compromising on reliability or measurement accuracy.
Straightforward Calibration Features Requires no dedicated adapter or setup for flow adjustment and calibration; compatible with all flow calibrators.
Advanced NiMH Battery and Enclosure Construction The Gilian GilAir-5 can operate for a minimum of 8 hours thanks to its advanced battery. The enclosed flow adjustment and on/off switch compartment protects the pump against dust, dirt, and moisture, preventing unauthorised tampering with the pump.
Consider GilAir-5 for IOM at 2 L/min and GK2.69 at 1.6 / 4.2 L/min, subject to filter loading. Low-flow gas/vapour sampling needs the appropriate module/holder. Set up the complete sampling train, verify flow at the sampling-head inlet before and after use, and record the result. A field flow check does not replace periodic laboratory calibration.

GK2.69
Thoracic fraction at 1.6 L/min; respirable fraction at 4.2 L/min. Select the fraction by flow and confirm that the pump maintains it under filter load.

IOM · 2 L/dk
Samples the inhalable dust fraction at 2 L/min. Select the filter and internal cassette for the weighing/analytical method.

25 / 37 mm · metot seçimi
Choose MCE, PVC, PTFE or another filter for the analyte and laboratory method. Match cassette diameter, sections, pore size and flow in the sampling plan.

Sorbent · metot seçimi
For gas/vapour sampling select sorbent, tube dimensions, duration and low flow by method. Tube holder, breaker and connection tubing complete the train.
Impinger · sıvı tutucu
Liquid collection methods require an upright bottle holder and a trap protecting the pump. The analytical method determines flow, absorbing liquid and sample volume.

Gilibrator 3 · hücre seçimi
Dry cells: low 5–450 mL/min, standard 50–5000 mL/min, high 1–30 L/min. Select the cell covering the sampling flow; check before/after sampling with head, filter and tubing assembled.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.
Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.