Customer need
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
NOVEL · Product and System Solutions
Occupational Safety and Environmental Measuring Instruments

From need to solution
Gilian LFS-113: selection starts with the measurement or system output your team needs for its decision.
Sampling requires air to be drawn from the breathing zone or a selected location at the flow and duration specified by the method.
Air passed through the appropriate filter, cassette, cyclone or sorbent tube creates a sample for analysis; the pump alone does not measure contaminant concentration.
Relate the laboratory result to sampled air volume, selecting media and flow together and calibrating the assembled sampling train.
The Gilian LFS-113 D is a personal air sampling pump used for low-flow applications such as sorbent tube and bag sampling. It operates at 20-200 cc/min in constant flow mode and 1-350 cc/min in constant pressure mode.
Select a suitable sorbent tube or sampling bag for the analytical method. Check the flow before and after sampling with an appropriate calibrator and the complete sampling train.
Only the LFS-113 DC Clock variant with a timer display has been discontinued; the basic model remains in active production.
| Flow range | Constant flow: 20–200 cc/min, to 25 in H₂O. Constant pressure: 1–350 cc/min, to 18 ±3 in H₂O. |
|---|---|
| Constant flow control | Better than ±5% of set flow after calibration. |
| Flow fault | Indicator; shutdown if a flow deviation over 5% lasts more than 30 seconds. |
| Battery | Rechargeable NiMH; 14–18 hour charge time. |
| Runtime | Manufacturer specification: minimum 8 hours; actual use depends on the sampling setup and battery condition. |
| Weight | 340 g |
| Approvals | Manufacturer datasheet lists CSA, CE and UL, including ATEX II 2 G EEx ib IIC T4. Confirm the certificate and approved configuration for the intended area. |
Specify the complete code in your quotation request. Kit contents, plug type and current supply options are confirmed separately.
LFS-113 D and LFS Pro are separate models. Check the pump model before ordering a charger or spare.
| Manufacturer Code | Accessory / Configuration |
|---|---|
| 810-0301-02 | LFS-113 D basic pump kit, no charger |
| 910-0301-02 | Starter kit with 230 V Euro-plug charger |
| 910-0302-02 | Five-pump kit with 230 V Euro-plug charger |
| 298-0005-01 | Single charger, US plug |
| 400198-1 | Single charger, Euro plug |
| 400324 | Five-unit charger, US plug |
| 400373 | Five-unit charger, Euro plug |
| 800685 | Air Boss bag sampling kit |
| 800400 | Carrying case, 45.7 × 33 × 17.8 cm |
Source: current manufacturer datasheet. View datasheet
LFS-113 is for low-flow gas/vapour sampling: 20–200 mL/min constant flow. Do not pair it with cyclone/IOM heads requiring 1.6–4.2 L/min. Set up the complete sampling train, verify flow at the sampling-head inlet before and after use, and record the result. A field flow check does not replace periodic laboratory calibration.

Sorbent · metot seçimi
For gas/vapour sampling select sorbent, tube dimensions, duration and low flow by method. Tube holder, breaker and connection tubing complete the train.
Impinger · sıvı tutucu
Liquid collection methods require an upright bottle holder and a trap protecting the pump. The analytical method determines flow, absorbing liquid and sample volume.

Gilibrator 3 · hücre seçimi
Dry cells: low 5–450 mL/min, standard 50–5000 mL/min, high 1–30 L/min. Select the cell covering the sampling flow; check before/after sampling with head, filter and tubing assembled.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
Describe the measurement or communication task, operating environment, required range or capacity and recording needs. NOVEL can help review the model and configuration information on this page.
Review the specifications, documents and any accessories sections on this page. Include model, quantity and intended use in your quotation request, and confirm compatibility before ordering.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.