Customer need
Noise control requires level and duration information appropriate to the assessment, beyond the perception of loudness alone.
NOVEL · Product and System Solutions
Occupational Safety and Environmental Measuring Instruments

TSI OmniWear Noise™ is a compact personal noise dosemeter developed to monitor and log the noise workers are exposed to throughout the working day. Thanks to Bluetooth® connectivity, instrument setup, live measurement tracking, remote start/stop, data transfer, and PDF report generation can be carried out via the TSI OmniWear Noise™ mobile app.
From need to solution
TSI OmniWear Noise™ Personal Noise Dosemeter: selection starts with the measurement or system output your team needs for its decision.
Noise control requires level and duration information appropriate to the assessment, beyond the perception of loudness alone.
Sound-level, exposure or analysis indicators supported by the chosen model are recorded with the field conditions.
Prioritise sources, further measurements and controls, distinguishing the role of a dosimeter from a sound-level meter and checking acoustic calibration requirements.
TSI OmniWear Noise™ – Product Overview
TSI OmniWear Noise™ – Use and Application
In the manufacturer's current technical documentation, the instrument is specified under ANSI S1.25:1991 R2024 and BS EN 61252:1997+A2:2017. ISO 9612 is not an instrument certification, but a measurement and calculation standard for employee noise exposure; therefore, the instrument standard, field calibration, and measurement methodology must be evaluated together.


Technical Specifications
| Application Type | Personal Noise Exposure Meter (Dosemeter) |
|---|---|
| Battery Type, Operating Life | Li-ion 350 mAh; typical 16 hours run time, 100% charge in approx. 6 hours |
| Data Logging | Maximum 64 measurement runs; 1-second time-history profile and data transfer / PDF reporting via mobile app |
| Instrument Kit Contents | OmniWear Noise unit, charger module, windscreen, protective accessories depending on configuration; 5/10/20-unit kit options |
| Weight | 36 g |
| Measurement Parameters | LXY, LXYmax, LXeq, LXpeak, LAvg, LXIeq, LTM3, LTM5, LAE; LEX,8h, TWA, % Dose, projected dose, Pa²h, Pa²s, motion %, LASmax |
| Standards | ANSI S1.25:1991 R2024; BS EN 61252:1997+A2:2017 |
| Response Range | Fast and Slow |
| Weighted Measurement | A, C, and Z; exchange rate Q=3 or Q=5 dB |
| Linear Operating Range | 70–140.3 dB(A) RMS |
| Peak Measurement Range | 90.0–143.3 dB(C or Z) |
| Protection Class | IP65 |
| Dimensions | 56 × 32 × 40 mm |
| Bluetooth / Mobile | Live monitoring, start/stop, and reporting via mobile app; control/monitoring of up to 20 instruments according to manufacturer announcement |
Complies with the IEC 61252 noise dosemeter standard; supports common measurement protocols such as ISO 9612, OSHA, ACGIH, and MSHA.
Wireless connection via Bluetooth® to a phone or tablet; remote monitoring and control of multiple dosemeters.
Supports A, C, and Z frequency weightings along with Fast and Slow time weightings.
1-second time-history profile, motion detection, pause function, and simultaneous noise dose parameters.
Robust IP65-rated enclosure, MEMS microphone, and approximately 16 hours of run time.
Modular wireless charging system; up to 10 charging modules can be linked together using a single power supply.
Measurement results can be viewed in the mobile app; automated PDF reports can be shared via phone or tablet.
Use the manufacturer-specified calibrator and matching microphone adapter for pre/post field checks. For Edge 7/8 use 114 dB at 1 kHz. Field checks do not replace periodic instrument/calibrator laboratory calibration.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.