NOVEL · Product and System Solutions

TSI Quest Edge 8 Noise Dosimeter

TSI Quest Edge 8 ATEX / IECEx Personal Noise Dosimeter

Occupational Safety and Environmental Measuring Instruments

TSI Quest Edge 8 ATEX IECEx personal noise dosimeter

TSI Quest Edge 8 ATEX / IECEx Personal Noise Dosimeter

The TSI Quest Edge 8 is an intrinsically safe personal noise dosimeter developed to safely monitor the personal noise exposure of workers in industrial environments with explosive atmosphere risks. In addition to ATEX and IECEx certification, it provides four independently configurable dosimeters, BLE5 Bluetooth®, 1/1 octave band analysis, voice notes, event audio recording, and the TSI Detection Management Software (DMS) workflow. It supports detailed assessment of personal noise exposure in mining, oil and gas, refineries, chemical processing, and other hazardous area applications.

Request a Quote from NOVEL ↗



ATEX / IECEx Intrinsically Safe Personal Noise Dosimeter

Sole Authorised TSI Distributor in Turkey: NOVEL Ltd.
You may contact NOVEL for product selection, sales, technical support, training, and after-sales services.
Product Specifications
  • Supports management of field studies via BLE5 Bluetooth and the Edge dB mobile application.

  • Helps evaluate different exposure criteria in the same study using four independently configurable virtual dosimeters.

  • Offers advanced analysis functions such as 1/1 octave band data logging, voice notes, and event audio recording.

  • Product Documents

      Technical Specifications

      Product Videos

      TSI Quest Edge 8 product video
      TSI Quest Edge 8 product video
      Watch Video ↗

      Technical Specifications

      Technical Table 1
      ModelEG8 / EG8-BT / EG8M-BT
      ConnectivityUSB via EdgeConnect docking; BLE 5 / Edge dB application on Bluetooth models
      Virtual Dosimeter4 independently configurable
      Frequency Analysis1/1 octave band analysis
      Microphone1/2-inch MEMS, field replaceable microphone
      Data ManagementEdge dB mobile application and TSI DMS
      Measurement Range70–140 dB RMS; 110–143 dB Peak
      Potentially Explosive AtmosphereATEX / IECEx intrinsically safe
      Ingress ProtectionIP65
      RMS Measurement Range70–140 dB
      Peak Measurement Range110–143 dB
      RMS Frequency WeightingA or C
      Peak WeightingC or Z
      Time WeightingFast or Slow
      Independent Dosimeter4 independent dosimeters
      Threshold70–90 dB in 1 dB steps or off
      Exchange RateIndependently 3 dB, 4 dB, or 5 dB for each dosimeter
      Criterion Level70–90 dB in 1 dB steps
      StandardsANSI S1.25-1991 (R2017); IEC 61252:1993 + AMD1:2000 + AMD2:2017
      ATEX / IECExATEX Directive 2014/34/EU; Ex ia IIC T4 Ga; Ex ia I Ma; IECEx UL24.0016; ATEX UL 24ATEX3178
      BatteryRechargeable Li-Polymer; nominal 30 hours with display off
      MemoryLogging capacity up to 100 days with 1-minute logging
      Protection ClassIP65
      Dimensions / Weight123.4 × 59.2 × 32 mm / 4.16 oz (approximately 118 g)
      Advanced FunctionsPause Study, Voice Notes, event audio recording, Ceiling Count, 1/1 octave analysis, and DMS reporting
      Technical Table 2
      İSGÜM / TS EN ISO 9612 ApproachIn İSGÜM's laboratory authorization resources, the TS EN ISO 9612 task-based, job-based, and full-day measurement approach is used for personal noise exposure. Job analysis, appropriate measurement strategy, measurement, uncertainty evaluation, and reporting must be conducted together.
      Personal Exposure MeasurementFor comparison with occupational exposure limit values, personal noise exposure must be monitored from a point close to the worker's hearing zone. The Edge 8 supports shift- and task-based measurements with its shoulder-worn personal dosimeter design.
      Four Independent DosimetersEdge 8 can run four independent dosimeter profiles simultaneously; threshold, exchange rate, and criterion level parameters can be independently configured according to corporate procedures or different assessment criteria.
      ATEX / IECEx Hazardous AreaTSI's current technical datasheet for the Edge 8 specifies ATEX Directive 2014/34/EU and IECEx certifications. The instrument supports Ex ia IIC T4 Ga and Ex ia I Ma classifications. Prior to use, the field zone classification, gas group, temperature class, and current certification scope must be independently verified.
      Measurement Data for ISO 9612RMS, dose, TWA, LAVG/LEQ, peak, maximum/minimum values, and time-history records can be used in personal exposure evaluation. 1/1 octave analysis provides additional data for evaluating the frequency composition of noise and noise control measures.
      Calibration and Field CheckPre- and post-measurement acoustic verification must be conducted using an appropriate calibrator and a valid field procedure. Manufacturer safety instructions and site procedures must also be observed for accessories and operations to be used in hazardous areas.
      Raw Data and RecordingEdge 8 provides a logging capacity of up to 100 days with a 1-minute logging interval. Measurement data can be stored, analyzed, and reported via EdgeConnect and DMS.
      Application in TurkeyUnder the TS EN ISO 9612 method, instrument standard compliance alone is not sufficient. Together with calibration, microphone location, task durations, uncertainty calculation, and laboratory procedures; suitability with the facility's hazardous area classification must also be verified in ATEX/IECEx areas.

      Measurement and Standards Approach

      The TSI Quest Edge 8 has been developed for hazardous area applications with ATEX / IECEx intrinsically safe certification alongside ANSI S1.25 and IEC 61252 personal noise dosimeter standards. used in Turkey for evaluating personal noise exposure of workers TS EN ISO 9612 is not an instrument certification, but a methodology standard defining measurement strategy and field execution. The Edge 8 combines personal exposure measurement, four independent dosimeter profiles, RMS/Peak, dose, TWA, 1/1 octave band analysis, raw data logging, and DMS reporting workflow with hazardous area requirements.

      Official source: İSGÜM – Laboratory Authorization and TS EN ISO 9612 occupational hygiene noise resources ↗

      Sectors

      Alternative Products

      You can examine product details and technical specifications by selecting an alternative suited to your requirements.

      Frequently Asked Questions About TSI Quest Edge 8

      Which noise dosimeter standards does the Edge 8 meet?
      TSI's current technical datasheet specifies ANSI S1.25-1991 (R2017) and IEC 61252:1993+AMD1:2000+AMD2:2017 personal noise dosimeter / personal sound exposure meter standards for the Edge 8.
      Is the Edge 8 ATEX and IECEx certified?
      Yes. TSI's current Edge 8 technical datasheet specifies ATEX Directive 2014/34/EU and IECEx certifications, Ex ia IIC T4 Ga and Ex ia I Ma markings, along with certificate numbers. Suitability for the operational zone and gas group must be confirmed against current certificates prior to purchase and field deployment.
      Can the Edge 8 be used in TS EN ISO 9612 personal noise measurements?
      The Edge 8 is a dosimeter designed for monitoring personal noise exposure on workers on a shift- and task-based basis. TS EN ISO 9612 is not an instrument certification, but a measurement strategy and calculation standard; in field practice, instrument standard compliance, calibration, microphone position, task durations, and laboratory procedures must be implemented together.
      How are Edge 8 data transferred and reported?
      Edge 8 measurement data can be transferred to TSI Detection Management Software (DMS) via USB through the EdgeConnect docking station. On Bluetooth models, BLE 5 and the Edge dB app support monitoring and control from a safe distance during field operations.

      Notice: The technical information, product specifications and images on this page are based on manufacturer publications. The manufacturer may change specifications without notice. For current specifications or assistance in selecting a suitable product for your application, please contact NOVEL at info@novel.com.tr .

      NOVEL technical support

      For your application,
      let us select the right solution together.

      Request a Quote from NOVEL ↗

      A draft opens in your email client with the subject header of the relevant product or category.