Customer need
Noise control requires level and duration information appropriate to the assessment, beyond the perception of loudness alone.
NOVEL · Product and System Solutions
Occupational Safety and Environmental Measuring Instruments

From need to solution
TSI Quest SoundPro SE-DL Sound Level Meters and Noise Analysers: selection starts with the measurement or system output your team needs for its decision.
Noise control requires level and duration information appropriate to the assessment, beyond the perception of loudness alone.
Sound-level, exposure or analysis indicators supported by the chosen model are recorded with the field conditions.
Prioritise sources, further measurements and controls, distinguishing the role of a dosimeter from a sound-level meter and checking acoustic calibration requirements.
On the SoundPro platform, SE models provide summary data logging, whereas DL models additionally offer time history / data logging. Octave analysis options are selected according to the model.
SoundPro is a platform geared towards occupational hygiene, environmental, and industrial applications that require detailed noise analysis beyond basic sound level measurement. Frequency analysis, time/frequency weightings, and the measurement protocol must be configured according to the applied methodology; instrument class and relevant standards compliance should be matched to project specifications.
| Series / Models | SoundPro SE and DL family; Class/Type 1 and Class/Type 2 configurations |
|---|---|
| Frequency Analysis | Real-time 1/1 octave or 1/1 + 1/3 octave analysis depending on the model |
| Data Logging | SE: summary data logging; DL: time history / data logging |
| DL Logging Interval | Selectable between 1 second and 60 minutes |
| Frequency / Time Weightings | A, C, Z; Fast, Slow, and Impulse |
| Software | Data transfer, analysis, and reporting with TSI Detection Management Software (DMS) |
| Standards Compliance | CE marked, IEC 61326-1, IEC 61672-1, ANSI S1-4, ANSI S1.43, IEC 61260, ANSI S1.11, IEC61010-1, IEC 60268-16 with speech intelligibility option |
| Operating Temperature and Humidity | Between -10°C and +50°C, Humidity: 10% to 90% RH |
| Measurement Parameters | SPL, Lmax, Lpik, Leq, Lavg, SEL, LN(L1-L99), TWA, Taktm, Dose |
| Total Measurement Range | 0 - 140 db |
| Total Dynamic Measurement Range | 120 db |
| Instrument Weight | 0.54kg (Including battery) |
| Extension Cable | 15m (Without signal loss) |
The TSI Quest SoundPro SE-DL is a modular noise analysis platform developed for professional sound level measurement and detailed frequency analysis. Supporting occupational hygiene, environmental noise, industrial acoustics, and long-term measurement studies thanks to Class/Type 1 and Class/Type 2 microphone options; simultaneous A, C, and Z frequency weightings; Fast, Slow, and Impulse time weightings; 1/1 and 1/3 octave real-time analysis options; data logging; and TSI Detection Management Software (DMS) workflow.
Easy Operation with User-Friendly Interface Unlike equivalent analyser instruments, SoundPro instruments feature a large LCD screen for convenient instrument control and an intuitive menu system for ease of operation. They are also fully compatible with DMS software, which enables straightforward management of all Quest instruments. Furthermore, 1/1 and 1/3 octave band graphic screens are provided for displaying octave band analyses. Quest SoundPro instruments are equipped with advanced programmability and level-triggered stop-start features.
SoundPro Outdoor Measuring System (OMS) SoundPro instruments can also perform long-term measurements outdoors for environmental monitoring and provide environmental survey analysis. The OMS system was developed for such applications. OMS is an advanced system designed to protect the instrument against adverse environmental influences (such as wind, rain, chemicals, particulate matter, and theft) while providing the necessary power for long-term operation. With the OMS system, the instrument can perform continuous measurements for more than 150 hours (1 week). (Additional batteries can be supplied for longer campaigns.) The OMS construction comprises stainless steel, ABS, and polymeric materials, and features a dedicated lock system for security alongside a purpose-designed case facilitating easy transport at the end of the operation.
Use the manufacturer-specified calibrator and matching microphone adapter for pre/post field checks. For Edge 7/8 use 114 dB at 1 kHz. Field checks do not replace periodic instrument/calibrator laboratory calibration.

AC-300
Pre/post measurement acoustic checks with a supported microphone and matching adapter. Select level, microphone type and corrections according to the instrument manual.
056-990 · ½ inç
AC-300 adapter for the supported ½-inch SE-400/SoundPro/SD-200 microphone configuration. Confirm SoundPro ¼-inch and 1-inch options separately.
You can examine product details and technical specifications by selecting an alternative suited to your requirements.
NOVEL technical support
A draft opens in your email client with the subject header of the relevant product or category.